Section 1 Legal information
Manufacturer: AppliTek NV/SA
Distributor: Hach Lange GmbH
The translation of the manual is approved by the manufacturer.
Section 2 Specifications
Specifications are subject to change without notice.
Table 1 General specifications
Specification Details
Dimensions (W x H x D) 460 × 688 × 340 mm (18.11 × 27.09 × 13.39 in.)
Enclosure Enclosure rating: IP44; indoor use only
Enclosure material: ABS, PMMA and coated steel
Weight 25 to 40 kg (55 to 88 lb) (based on analyzer model)
Power requirements 110–240 VAC ±10%, 50/60 Hz
1
Power consumption 150 VA maximum
1
Installation category II
Pollution degree 2
Operating temperature 10 to 30 °C (50 to 86 °F); 5 to 95% relative humidity, non-condensing, non-corrosive
Storage temperature -20 to 60 °C (-4 to 140 °F), ≤ 95% relative humidity, non-condensing
Instrument air feed Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Minimum pressure: 6 bar (600 kPa or 87 PSI)
Demineralized water For rinsing and/or dilution
Drain Atmospheric pressure, vented, minimum Ø 64 mm
Earth connection Dry and clean grounding pole with low impedance (< 1 Ω) with an earth cable of >
2.5 mm
2
(13 AWG)
Analog inputs Electrodes, temperature, conductivity, colorimeter
Analog outputs Two to four 4–20 mA; maximum load: 500 Ω, galvanically isolated
2
Digital inputs Four digital inputs: remote start/stop (potential-free contact) (optional)
Digital outputs Four potential-free digital outputs (FCT) to control internal valves/pumps; 24 VDC
Four powered digital outputs to control external vavles/pumps; 24 VDC, 500 mA
1
Power requirements and consumption are based on the analyzer model, refer to the analyzer
serial number plate for detailed information.
2
Optional modules are available to add a maximum of 10 analog outputs to the analyzer.
English 3