surface of the water, the meter sets the velocity value to zero. For a profile or real-time reading, the
meter prompts the user to submerge the sensor in the water.
If the specific conductivity of the water being measured is very low, adjust the threshold value for the
best performance. Go to Setup Menu>Wet/Dry Threshold.
For troubleshooting, the present reading is used as a guide to set a custom threshold. Get a reading
in the water (wet) and then out of the water (dry). The threshold value must be between the wet and
dry value. For example, if the actual wet value is 17 and the actual dry value is 2, put in a threshold
value half way between 2 and 17. The default value is 20.
English 35