CAT. NO. 46500-88
PORTABLE TURBIDIMETER
Model 2100P
Instrument and Procedure Manual
© Hach Company, 1991-2004. te/dk 3/04 8ed
All rights reserved. Printed in the U.S.A.

Table of Contents

Questions and Answers

  • A
    Aaron MasseyAug 13, 2025
    What to do if Hach Measuring Instruments shows error E7 (light leak)?
    • M
      Mrs. Tina Rodriguez MDAug 13, 2025
      If the Hach Measuring Instruments displays error code E7, it indicates a light leak. Ensure the lid is closed properly before pressing the READ key to take a measurement.
  • R
    Richard ShawAug 16, 2025
    What to do if Hach 2100P Measuring Instruments shows error E3 (low light)?
    • M
      Michelle LeAug 17, 2025
      If the Hach Measuring Instruments displays error code E3, which indicates a low light error, try re-reading the measurement. Also, inspect the lamp and check for any obstructions in the light path. Dilution of the sample might also be necessary.
  • S
    Sarah AndersonAug 21, 2025
    What to do if Hach 2100P shows error E6 (A/D underrange)?
    • J
      Jessica DaughertyAug 21, 2025
      If the Hach Measuring Instruments displays error code E6, which indicates an A/D underrange, first ensure that the lid was not open during the reading, and then re-read the sample. Also, check for any obstructions in the light path. If the problem persists, contact Hach Service.
  • M
    Michael WashingtonAug 25, 2025
    What to do if Hach 2100P shows error E5 (A/D overrange)?
    • B
      Beth KramerAug 26, 2025
      If the Hach Measuring Instruments displays error code E5, which indicates an A/D overrange, check for any obstructions in the light path and call Hach Service.
  • J
    Justin CarterAug 28, 2025
    How to fix Hach Measuring Instruments error E8 (bad lamp circuit)?
    • A
      Amanda TuckerAug 28, 2025
      If the Hach Measuring Instruments displays error code E8, which indicates a bad lamp circuit, try reinserting the lamp leads at the terminal block, ensuring that the lead ends are not touching each other. If the error persists, contact Hach Service.
  • E
    Ellen NealSep 1, 2025
    What to do if Hach 2100P shows error E4 (EEPROM malfunction)?
    • S
      stevensonsusanSep 1, 2025
      If the Hach Measuring Instruments displays error E4, indicating an EEPROM malfunction, press the I/O button. If E4 reappears, call Hach service. If CAL? appears, recalibrate the instrument.
  • T
    Tara ReedSep 4, 2025
    Why does my Hach Measuring Instruments show error E2, and how can I fix it?
    • H
      halerobertSep 5, 2025
      Error E2 on the Hach Measuring Instruments indicates that two standards have the same value or their difference is less than 60 NTU, not all standards were read during the calibration, or standard 1 is too low (
  • R
    Randy MeyerSep 8, 2025
    How to fix Hach 2100P error E1 related to dilution water?
    • B
      Brendan KellySep 8, 2025
      If the Hach Measuring Instruments displays error code E1, it means the dilution water is ? 0.5 NTU. You should start the calibration over with better quality dilution water. Alternatively, use a membrane filter to filter the water before use.

Summary

CERTIFICATION

Product Safety, Immunity, and Emissions Compliance

Details safety standards (UL, CSA, VDE, GS, CE) and immunity/emissions compliance (EN, IEC, FCC, Canadian).

SAFETY PRECAUTIONS

Hazard Information and Precautionary Labels

Explains hazard signal words (Danger, Caution, Note) and safety symbols for user awareness.

SPECIFICATIONS

Performance Characteristics

Covers measurement method, range, accuracy, resolution, repeatability, response time, and stray light.

Instrument Components and Requirements

Details display, light source, detectors, signal averaging, sample cells, power, and battery life.

Operating and Storage Conditions

Specifies operating temperature, humidity range, dimensions, and weight.

INSTRUMENT DESCRIPTION

General Overview and Principle of Operation

Describes the turbidimeter's function, technology, and basic operation.

Preparation for Use

Covers unpacking, battery installation, and using power options.

TURBIDITY MEASUREMENT

Measurement Procedure and Techniques

Explains controls, indicators, step-by-step measurement, and essential techniques for accuracy.

Measurement Notes and Sample Handling

Provides important tips for accurate measurements, including cell care, degassing, and representative sampling.

INSTRUMENT OPERATION

Controls, Indicators, and Key Functions

Details the function of each key and display icon for operating the instrument.

Measurement Modes and Calibration Procedures

Explains range selection, signal averaging, and detailed calibration using primary and secondary standards.

MAINTENANCE

Cleaning and Battery Replacement

Instructions for keeping the instrument clean and replacing batteries.

Lamp Replacement Procedure

Step-by-step guide for replacing the instrument's lamp.

TROUBLESHOOTING

Diagnostic Functions and Procedures

Explains how to access and use diagnostic codes and procedures for troubleshooting.

Error Messages and Corrective Actions

Lists common error messages, their probable causes, and corrective actions.

GENERAL INFORMATION AND SUPPORT

Parts, Ordering, and Service Information

Information on replacement parts, accessories, how to order, and repair service details.

Hach 2100P Specifications

General IconGeneral
Range0 to 1000 NTU
Light SourceTungsten filament lamp
DetectorSilicon photodiode
Operating Temperature0 to 50 °C (32 to 122 °F)
Power4 AA alkaline batteries
Resolution0.01 NTU
Measurement PrincipleNephelometric
DisplayLCD