EasyManuals Logo

Hach 2100P User Manual

Hach 2100P
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
54
SECTION 3, continued
NOTES
If the I/O key is pressed during calibration, the new calibration data
is lost and the old calibration will be used for measurements. Once
in calibration mode, only the
READ, I/O, , and keys function.
Signal averaging and range mode must be selected before entering
the calibration mode.
If E1 or E2 are displayed, an error occurred during calibration.
Check the standard preparation and review the calibration; repeat
the calibration if necessary. If the error messages recur, calibrate
using the factory specified standards, Section 3.6.2.4 on page 42
and Section 3.6.3 on page 44. Press
DIAG to cancel the error
message (
E1 or E2). To continue without repeating the calibration,
press
I/O twice to restore the previous calibration. If CAL? is
displayed, an error may have occurred during calibration. The
previous calibration may not be restored. Either recalibrate or use
the calibration as is.
To review a calibration, press CAL and then only to view the
calibration standard values. As long as
READ is never pressed and
CAL isn’t flashing, the calibration will not be updated. Press CAL
again to return to the measurement mode.
3.6.4 Using Gelex
®
Secondary Turbidity Standards
Note: Store Gelex standards at room temperature. Do not allow to freeze or
exceed 50 °C.
The instrument comes with Gelex Secondary Standards which are
particulate suspensions similar to formazin primary standards in light
scattering characteristics. NTU values on the Gelex standards indicate
the range for which they should be used. Due to minor variations in
glass and individual instrument optical systems, the true value of the
Gelex standards must be determined against formazin in the same
instrument they will be used with for later calibration checks.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 2100P and is the answer not in the manual?

Hach 2100P Specifications

General IconGeneral
Range0 to 1000 NTU
Light SourceTungsten filament lamp
DetectorSilicon photodiode
Operating Temperature0 to 50 °C (32 to 122 °F)
Power4 AA alkaline batteries
Resolution0.01 NTU
Measurement PrincipleNephelometric
DisplayLCD

Related product manuals