EasyManuals Logo

Hach 2100P User Manual

Hach 2100P
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
37
SECTION 3, continued
3.6 Calibration
Calibration of the 2100P Turbidimeter is based on formazin, the
primary standard for turbidity. The instrument's electronic and optical
design provide long-term stability and minimize the need for frequent
calibration. The two-detector ratioing system compensates for most
fluctuations in lamp output. A formazin recalibration should be
performed at least once every three months, more often if experience
indicates the need. When calibration is necessary, use a primary
standard such as StablCal™ Stabilized Standards or formazin
standards.
Hach Company only recommends the use of StablCal
®
Stabilized
Formazin or formazin standards for the calibration of Hach
turbidimeters. Hach Company cannot guarantee the performance
of the turbidimeter if calibrated with co-polymer styrene
divinylbenzene beads or other suspensions.
Important Note: DO NOT calibrate with Gelex® Secondary Standards.
Gelex standards are designed for instrument verification, not calibration.
3.6.1 StablCal Stabilized Formazin Standards*
Most consistent results will be achieved with the use of StablCal
Stabilized Formazin Standards for calibration. Refer to Section 3.6.1.2
and Section 3.6.1.3 for information on preparing the standards for use.
Note: Hach StablCal Stabilized Formazin in 20-, 100-, and 800-NTU values is
packaged in convenient sets for calibration of the 2100P Turbidimeter. The set
may be ordered in 500-mL size bottles by specifying Cat. No. 26594-00, in
100-mL size bottles by specifying Cat. No. 26594-10 or in sealed vials by
ordering Cat. No. 26594-05. (See OPTIONAL ACCESSORIES AND
REAGENTS on page 74.)
3.6.1.1 Storing and Handling StablCal Stabilized Standards
For optimum results when using StablCal Stabilized Standards, adhere
to the following recommendations:
* StablCal Stabilized Formazin is cited as a primary standard in Hach Method
8195, an acceptable version of USEPA Method 180.1.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 2100P and is the answer not in the manual?

Hach 2100P Specifications

General IconGeneral
Range0 to 1000 NTU
Light SourceTungsten filament lamp
DetectorSilicon photodiode
Operating Temperature0 to 50 °C (32 to 122 °F)
Power4 AA alkaline batteries
Resolution0.01 NTU
Measurement PrincipleNephelometric
DisplayLCD

Related product manuals