32
SECTION 2, continued
Use secondary standards for periodic calibration checks. Please note
that Gelex
®
standards must be assigned values after StablCal Stabilized
Formazin calibration or formazin calibration and before use as
secondary standards. Gelex standards must be recalibrated (values
assigned) each time the instrument is calibrated with StablCal
Stabilized Formazin or formazin. See Section 3.6 on page 37 for
detailed information on the use of StablCal Stabilized Formazin,
formazin, and Gelex standards.
2.3.9 Representative Sampling
A representative sample accurately reflects the true condition of
the water source from which the sample was taken. To ensure a
representative sample, gently, but thoroughly, mix every sample
before aliquots are taken. Do not allow the sample to settle.
When sampling from a tap in a distribution system or treatment plant,
allow the water to run for at least five minutes before sampling.
When sampling from a stream, reservoir, clarifier, or storage tank,
collect at least one liter (l quart) and thoroughly mix before
measurement. If the water source is not uniform, it may be necessary
to sample several locations at varying depths and combine the samples
into a single, well-mixed composite sample before measurement.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 2100P and is the answer not in the manual?

Hach 2100P Specifications

General IconGeneral
Range0 to 1000 NTU
Light SourceTungsten filament lamp
DetectorSilicon photodiode
Operating Temperature0 to 50 °C (32 to 122 °F)
Power4 AA alkaline batteries
Resolution0.01 NTU
Measurement PrincipleNephelometric
DisplayLCD

Related product manuals