SECTION 2, continued
2.2.2 Measurement Notes
Always cap the sample cell to prevent spillage of sample into
the instrument.
When taking a reading, place the instrument on a level, stationary
surface. It should not be held in the hand during measurement.
Always close the sample compartment lid during measurement
and storage.
Always use clean sample cells in good condition. Dirty, scratched,
or damaged cells can cause inaccurate readings.
Do not leave a sample cell in the cell compartment for extended
periods of time. This may compress the spring in the cell holder.
Remove sample cell and batteries from instrument if the instrument
is stored for extended time period (more than a month).
Avoid operating in direct sunlight.
Make certain cold samples do not “fog” the sample cell.
Avoid settling of sample prior to measurement.
Keep sample compartment lid closed to prevent dust and dirt
from entering.
2.3 Measurement Techniques
Proper measurement techniques are important in minimizing
the effects of instrument variation, stray light and air bubbles.
Regardless of the instrument used, measurements are more accurate,
precise and repeatable if the analyst pays close attention to proper
measurement techniques.
Measure samples immediately to prevent temperature changes and
settling. Avoid sample dilution when possible. Particles suspended in
the original sample may dissolve or otherwise change characteristics
when the sample temperature changes or when the sample is diluted,
resulting in a non-representative sample measurement.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 2100P and is the answer not in the manual?

Hach 2100P Specifications

General IconGeneral
Range0 to 1000 NTU
Light SourceTungsten filament lamp
DetectorSilicon photodiode
Operating Temperature0 to 50 °C (32 to 122 °F)
Power4 AA alkaline batteries
Resolution0.01 NTU
Measurement PrincipleNephelometric
DisplayLCD

Related product manuals