Error/Warning Description Solution
Meter will not power
on or powers on
intermittently
Batteries are not
installed correctly
Examine battery orientation to
make sure the batteries follow
the polarity markings. Test again.
Clean the battery terminals, then
install new batteries.
Connect AC power adapter and
test again.
Software not
updated to most
current version
To download the most current
version of the software, refer to
the applicable product page on
the manufacturer's website.
Damaged meter Contact Technical Support.
Unable to access
Full Access Options
screen
Correct password
has not been
entered
Contact Technical Support.
Unable to access
Full or Operator
Access Options
screen
Software not
updated to most
current version
To download the most current
version of the software, refer to
the applicable product page on
the manufacturer's website.
English 17

Other manuals for Hach HQ40d

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach HQ40d and is the answer not in the manual?

Hach HQ40d Specifications

General IconGeneral
Resolution0.1 °C
ConnectivityUSB
Storage Temperature-20 to 60 °C
Waterproof RatingIP67
DisplayLCD with backlight
Measurement ParameterspH, Conductivity, Dissolved Oxygen, ORP
mV Range-1999 to 1999 mV
Conductivity Range0.01 µS/cm to 200 mS/cm
Temperature Range-10 to 110 °C
Data Logging500 data points
Power Supply4 AA batteries
Accuracy±0.1 °C
Operating Temperature0 - 50 °C (32 - 122 °F)