    Duane WoodSep 12, 2025
    What does E42 error mean on Hach H160G?
    • L
      lisahicksSep 12, 2025
      Error E42 indicates invalid input. The value entered during the setup is invalid. Enter a different value.
  • B
    brandirandallAug 22, 2025
    What to do if my Hach H160G Measuring Instruments show an unstable reading?
    • J
      Jocelyn HoodAug 23, 2025
      If you are experiencing unstable readings with your Hach Measuring Instruments, several factors could be the cause. First, the probe might be dirty, so try cleaning it. Also, make sure to clean the probe contacts on the cable connector and the meter using methanol on a cotton swab, letting it dry completely before reconnecting. Other potential issues include no flow in the reference junction (clean the warm buffer), an ISFET probe that isn't correctly hydrated (soak it in pH 7.00 buffer for at least 5 minutes), interference from other devices (remove them), or direct sunlight (shield the probe). Note that stable readings may not be possible in very low ionic strength solutions or if the pH or temperature of the solution is changing.
  • S
    Scott GriffinAug 25, 2025
    What does E14 error mean on Hach H160G?
    • T
      Tyler MartinAug 25, 2025
      Error E14 on your Hach Measuring Instruments indicates that the battery is extremely low. Replace the batteries immediately to prevent potential damage to the accuracy and function of the meter.
  • S
    Samantha ThomasAug 27, 2025
    How to fix a Hach H160G with no display?
    • P
      Preston HintonAug 28, 2025
      If your Hach Measuring Instruments have no display, it could be due to a couple of reasons. First, check if the automatic shutdown feature has turned the power off; if so, simply set the power back to on. If that's not the issue, the problem might be a lack of power, in which case you should replace the batteries.
  • L
    Lisa WilsonAug 29, 2025
    What does E04 error mean on Hach H160G Measuring Instruments?
    • T
      Thomas HendricksAug 29, 2025
      Error E04 indicates a glass pH probe slope error, meaning the slope is smaller than 85% or larger than 102% of 59.16 mV per pH unit. Clean the probe. If the error continues, replace the probe.
  • D
    dannymillerAug 31, 2025
    What does E06 error mean on Hach H160G?
    • B
      bonniemcconnellAug 31, 2025
      Error E06 indicates an ISFET pH probe slope error. Clean the probe, soak it in 40 ºC (113 ºF) pH 7.00 buffer for 2 minutes, and calibrate again. If the error continues, replace the probe.
  • A
    Anna StricklandNov 25, 2025
    Why does my Hach Measuring Instruments meter continually show –2.00 or 19.99?
    • J
      Jordan TranNov 25, 2025
      If your Hach Measuring Instruments meter continually shows –2.00 or 19.99 with an electrode attached, and the "ISFET" icon does not appear when an ISFET probe is attached, it could be due to several reasons. First, the ISFET probe may not be sensed by the meter. Try setting the meter power to off and on again. If the "ISFET" icon still does not show, the ISFET probe may need replacement. Other potential causes include the probe being out of calibration (calibrate the ISFET probe), not being in the solution (put the probe in liquid, ensuring no air bubbles are caught on the sensor surface), not being connected properly (set the meter power to off, connect the pH probe, and set the meter power to on), or being dirty (clean the probe). It is also possible that the probe is damaged, requiring ...
  • B
    Brittany PhelpsSep 13, 2025
    What does E44 error mean on Hach H160G Measuring Instruments?
    • N
      Nicole NunezSep 13, 2025
      Error E44 indicates that no probe is installed for the applicable parameter. Set the meter power to off. Install the correct probe. Set the meter power to on.
  • B
    Brenda RodriguezSep 10, 2025
    What does E40 error mean on Hach Measuring Instruments?
    • V
      Vanessa SmithSep 10, 2025
      Error E40 indicates an unrecognized host command. This means a command from a host PC is not found. Use only valid commands.
  • D
    Donald WellsSep 6, 2025
    What does E27 error mean on Hach Measuring Instruments?
    • T
      thompsonkaraSep 6, 2025
      Error E27 indicates a DO probe temperature error. Attach the temperature sensor to the 3.5 mm phono jack. DO readings are highly temperature dependent, and a temperature sensor must be attached.