Set the data log
Note: The data log procedure stops during calibration and setup. Calibration alarms are not active during the data
log procedure.
1. When the meter is in read mode, push LOG to record data. The "LOGGING" icon flashes and the
digits in the lower right corner shows the number of collected data points.
2. To stop a data log procedure, push LOG.
3. To continue a data log procedure, push LOG again.
Note: Monitor data log memory to prevent data loss. If the memory is full, the meter deletes every even-
numbered data point and continues to record data.
Recall the data
To recall data in storage or logged data:
1. In the measurement mode, push RECALL.
2. Push STORE to recall data in storage or push LOG to recall logged data.
3. To see the last data storage location, push RECALL.
4. To select a different location, use the numeric keypad to enter the location number, then push
ENTER.
Note:
If the user tries to recall data from a location with no data in storage, the display shows "NO." If "NO"
shows, use the arrow keys to scroll to another data storage location.
5. To show the time a sample in storage was recorded, push TIME.
6. To show the date a sample in storage was recorded, push TIME again.
7. Push pH, ISE, COND, TDS, SAL or DO or CANCEL to go back to measurement mode.
Erase a single data measurement
Erase a data measurement or specific data measurements. Refer to Erase all measurement data
on page 17 to erase all data measurements.
1. Push RECALL in the measurement mode.
2. Push STORE CAL or LOG to recall calibration or logged data records. The number of the last
data measurement location shows.
3. Push RECALL to see the data.
4. To select a different memory location, use the numeric keypad to enter the location number, then
push ENTER.
5. When the data to delete shows on the display, push DELETE. The meter will show "DELETE ?".
6. Push ENTER to erase a data measurement.
Note: To exit and not keep the changes, push CANCEL.
7. Push pH, ISE, COND, TDS, SAL or DO to go back to the measurement mode.
Erase all measurement data
Refer to Erase a single data measurement on page 17 to erase specific data measurements. To
erase all data in storage or all logged data measurements:
1. With a sample shown, push DELETE. The meter shows "DELETE SAMPLE?".
2. Push STORE or LOG to recall data. The number of the last data measurement location shows.
3. Push RECALL to show the data.
4. Push 0, then ENTER and the meter shows "ALL."
5. Push ENTER to erase all data in storage or logged data.
Note: To exit and not keep the changes, push CANCEL.
Keep calibration data
The meter keeps the last 10 calibrations. Table 2 shows the date for each calibration.
English 17