Option Function Description
date Date format Sets the date to mm/dd/yy or dd/mm/yy format.
m/d date Month and day Use the number keys to set the correct month and day.
time Time Sets the correct time in a 24-hour format. Use the number keys to set the
correct time.
oFF Automatic
shutdown
Sets the shutdown parameters from 1 minute to 000 minutes (continuous
power). The meter beeps 1 minute before shutdown. Make sure to power
cycle for the automatic shutdown to occur. Automatic shutdown is disabled
during: data logging, Bluetooth
™
transfers and when connected to the USB
port.
Snd Sound options Sets sound alerts to on or off. Three sounds tell the user about different
functions:
• One beep: Stabilization occurred with the stabilization lock on.
• Two beeps: An error occurred. The error code shows on the display.
Refer to Troubleshooting on page 28.
• Three beeps: Measurement stability in calibration mode, regardless of
the stabilization lock setting.
ºC ºF Temperature
display
Sets the temperature unit from ºC to ºF.
READY Stabilization lock Sets the stabilization lock to on or off:
• The "READY" icon shows when an endpoint occurs and the stabilization
lock is set to on.
• After stabilization, the display locks the value and the meter ignores slight
measurement changes in pH, conductivity or TDS.
• Sets the stabilization lock to off during titrations or slight change
detection. The display automatically unlocks after a significant
measurement change is found.
1
This does not erase calibration data for an ISFET probe.
4. Push ENTER to keep the changes and go back to setup mode.
Note: To exit and not keep the changes, push CANCEL.
Operation
Configure the Bluetooth
™
wireless connection
Transfer data and control several instrument functions from a remote location to a PC with the
Bluetooth
™
wireless connection.
Note: The arrow icon shows on the display when the meter is in communication with the SmartLogger II software
on a PC. The arrow icon flashes when data is transferred. Refer to the SmartLogger II documentation to setup the
wireless connection to a PC.
1. Set the power to on. The meter looks for the Bluetooth
™
connection with a PC.
2. On the PC, select the option to find or add a Bluetooth
™
device. Next, the PC prompts if the user
wants to pair with a found device.
3. When the PC prompts for a pass key or PIN, enter the PIN (default = 1234). The PC shows if the
pair is successful and the meter shows the Bluetooth
™
icon.
4. If the pair is not successful, do steps
1 through 4 again.
Note: If there is more than one meter in range, each meter is identified with the model number and the serial
number (e.g., H170G LP SN1 2755).
English 15