What to do if Hach Lange DR 3900 test program stopped and asks to remove the cuvette?
J
Jenna FisherAug 30, 2025
If the test program stops on your Hach Measuring Instruments and prompts you to remove the cuvette and close the lid, it means the test program stopped when the instrument was started. Remove the cuvette/sample cell from the cell compartment and press OK.
C
Christopher CookAug 29, 2025
What to do if Hach Lange DR 3900 says 'USB memory stick is not connected'?
C
Carla BrownAug 30, 2025
If your Hach Measuring Instruments displays that the USB memory stick is not connected, it means an update is not possible. Check the USB memory stick to resolve this.
B
bethanyhebertAug 29, 2025
Why can't Hach Lange DR 3900 reach the remote server?
T
Todd AndersonAug 29, 2025
If the remote server cannot be reached on your Hach Measuring Instruments, it indicates an error during network setup. Ensure that the instrument is connected to the network.
B
Brianna WillisAug 28, 2025
How to fix 'Error Too much ambient light!' on Hach Measuring Instruments?
A
Alexander KimAug 28, 2025
If your Hach Measuring Instruments displays 'Error Too much ambient light!', it means the instrument sensors are detecting too much ambient light. Decrease the ambient light by avoiding direct sunlight or close the lid.
J
jessicacruzAug 27, 2025
What does 'Under measuring range' mean on Hach Measuring Instruments?
A
Alexis RitterAug 27, 2025
If your Hach Measuring Instruments indicates that you are under the measuring range, it means the measured absorbance is below the calibration range of the test. If possible, select a test with a lower reading range or use a cuvette with a longer path length.
K
Kathy BrownAug 27, 2025
What does 'Fault Clean cuvette!' mean on Hach Lange DR 3900 Measuring Instruments?
N
Nancy FloresAug 27, 2025
If your Hach Measuring Instruments displays 'Fault Clean cuvette!', it indicates that the cuvette is dirty or there are undissolved particles in the cuvette. You should clean the cuvette and allow any particles to settle.
M
Meghan EllisAug 26, 2025
What to do if Hach Lange DR 3900 doesn't read the barcode label?
C
cannonerinAug 26, 2025
If the barcode label is not read by your Hach Measuring Instruments, it could be due to a faulty barcode. Try reinserting the cell. If the barcode continues to not be recognized, contact the manufacturer or their representative.
Z
Zachary McfarlandAug 25, 2025
What to do if Hach Lange DR 3900 shows 'Concentration too high'?
J
Joshua BrownAug 26, 2025
If your Hach Measuring Instruments displays 'Concentration too high!', it means the calculated concentration is higher than 999999. To fix this, dilute the sample and measure it again.
M
melissa30Aug 25, 2025
What does 'Over measuring range' mean on Hach Measuring Instruments?
M
Maria GarzaAug 25, 2025
If your Hach Measuring Instruments shows that you are over the measuring range, it means the measured absorbance is above the calibration range of the test. Dilute the sample and measure again to resolve this.