What to do if absorbance is greater than 3.5 on Hach Lange DR 3900?
M
Michelle PorterAug 25, 2025
If the absorbance measured by your Hach Measuring Instruments exceeds 3.5, it means the measured absorbance is too high. To resolve this, dilute the sample and measure it again.
E
Emily BanksSep 1, 2025
What to do if Hach Lange DR 3900 Measuring Instruments says 'File for instrument update missing'?
C
christiannancySep 1, 2025
If your Hach Measuring Instruments displays that the file for the instrument update is missing, it indicates an error during the update process. Check the USB memory stick.
J
James PhillipsSep 1, 2025
Why am I getting an 'Error in FTP connection' on my Hach Lange DR 3900?
E
elizabeth79Sep 2, 2025
If there is an error in the FTP connection on your Hach Measuring Instruments, it indicates an FTP error. Make sure that the instrument is connected to the network.
B
Benjamin EstradaSep 2, 2025
Why is Hach Lange DR 3900 recommending a Full System Check?
M
Mark MartinSep 3, 2025
If it’s recommended to execute a Full System Check on your Hach Measuring Instruments, it means the check of the air values failed. Switch the instrument off and then back on again. If the system check is not successful, contact the manufacturer or his representative.
K
Katelyn RosalesSep 3, 2025
Why is the network switched off on my Hach Lange DR 3900 Measuring Instruments?
T
Tyler PerrySep 3, 2025
If the network is switched off on your Hach Measuring Instruments, it means the network setup is off when accessing the Instruments homepage via the sidebar. Activate the online connection.
S
Sherri HowardSep 4, 2025
Why does Hach Lange DR 3900 Measuring Instruments say 'Attention! Please insert the light shield'?
M
Mrs. Erin ColemanSep 4, 2025
If your Hach Measuring Instruments prompts you to insert the light shield, it is because readings with barcode cuvettes generally require the light shield. Insert the light shield and press OK.
M
mcleanmatthewSep 5, 2025
What to do if Hach Lange DR 3900 Measuring Instruments says 'Entry invalid!'?
R
Randy ButlerSep 5, 2025
If your Hach Measuring Instruments displays 'Entry invalid!', it means the password entered is incorrect. If you've forgotten your password, contact the manufacturer or their representative.
S
Sarah HughesSep 5, 2025
What does 'Fault Barcode check number? Please update program data!' mean on Hach Measuring Instruments?
J
Joshua OwensSep 6, 2025
If your Hach Measuring Instruments displays 'Fault Barcode check number? Please update program data!', it indicates a deviation to the stored data. Update the program data.
B
bentleychrisSep 6, 2025
What to do if Hach Lange DR 3900 Measuring Instruments says 'File for instrument update is faulty'?
K
kennethhenrySep 6, 2025
If your Hach Measuring Instruments displays that the file for the instrument update is faulty, it indicates an error during update. Save the update file again and repeat the procedure.