EasyManuals Logo

Hach LPV445.99.02110 User Manual

Hach LPV445.99.02110
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
7. Remove the blank sample cell.
8. Prepare the sample. Refer to the test procedure.
9. Clean the sample cell with a no-lint cloth.
10. Insert the sample cell into the cell holder. Make sure to install the
sample cell in the correct and consistent orientation so that the
results are more repeatable and precise. Refer to Figure 5.
11. Install the instrument cap over the cell holder. Refer to Figure 6.
12. Push . The display shows the results in concentration units or
absorbance.
Note: The result flashes if the result is less or more than the instrument
range.
13. Remove the sample cell from the cell holder.
14. Immediately empty and rinse the sample cell. Rinse the sample cell
and cap three times with deionized water (or distilled water).
Note: As an alternative, use tap water to rinse the sample cell if the samples
measured have a higher concentration than the tap water.
12 English

Other manuals for Hach LPV445.99.02110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach LPV445.99.02110 and is the answer not in the manual?

Hach LPV445.99.02110 Specifications

General IconGeneral
BrandHach
ModelLPV445.99.02110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals