Questions and Answers

  • R
    rchangAug 20, 2025
    Why does Hach Measuring Instruments show error E-01?
    • W
      William BrownAug 20, 2025
      Error E-01 on Hach Measuring Instruments indicates that the system detects too much ambient light. Ensure the instrument cap is fully installed over the cell holder. You can also check the service log for other error messages, inspect the cell compartment for ambient light sources, check the instrument cap for defects, and verify the light detection system is working. If needed, replace the cell compartment or detector.
  • L
    lbrownAug 22, 2025
    What to do if Hach Measuring Instruments show error E-02?
    • W
      wardjenniferAug 22, 2025
      If your Hach Measuring Instruments display error E-02, it means the LED light source control on the mainboard is out of regulation. Try replacing the batteries. Also, make sure the LED in the cell holder comes on when 'read' or 'zero' is pushed. Check the service log for other error messages and inspect the LED for damage. Furthermore, check the pin contact of the LED to the mainboard.
  • B
    Brandy SmithAug 25, 2025
    What to do if Hach LPV445.99.09110 Measuring Instruments show error E-20?
    • R
      Ryan SmithAug 25, 2025
      If your Hach Measuring Instruments display error E-20, this indicates too much signal on the light detector. Ensure the instrument cap is fully installed on the cell holder. Additionally, check the absorbance reading in 'service mode' and the absorbance values for service test kit vials. Also, check the optical aperture in front of the LED is still in place. If necessary, replace the optical aperture, optical filter, LED, or detector.
  • D
    Dennis BrownAug 26, 2025
    What to do if Hach LPV445.99.09110 show error E-12?
    • K
      Kelly GarciaAug 26, 2025
      If your Hach Measuring Instruments display error E-12, it means the battery power is too low. Replace the batteries. Also, check the battery power and ensure the battery contacts have appropriate contact with the mainboard. You may need to replace the battery contacts.
  • J
    justinhillAug 28, 2025
    What to do if Hach Measuring Instruments show error E-15?
    • S
      Shannon WareAug 28, 2025
      If your Hach Measuring Instruments display error E-15, it means the absorbance was read > 3.5 Abs. Check for any light blockage in the cell holder and clean the cell holder. If needed, dilute the sample. You can also check the absorbance reading in 'service mode' and the absorbance values for service test kit vials. Consider replacing the detector, LED, or optical filter.
  • A
    allisonwilliamsAug 29, 2025
    What to do if Hach LPV445.99.09110 show error E-21?
    • T
      Tiffany McfarlandAug 29, 2025
      If your Hach Measuring Instruments display error E-21, it indicates an unstable signal on the light detector. Make sure that the instrument cap is fully installed on the cell holder. Check the absorbance reading in 'service mode'. Perform a test measurement with any vial. If the issue persists, consider replacing the detector or LED.
  • C
    cindyharveyAug 31, 2025
    What to do if Hach LPV445.99.09110 Measuring Instruments show error E-04?
    • Z
      zhowellAug 31, 2025
      If Hach Measuring Instruments show error E-04, and the reading is less than the instrument range, make sure that the instrument cap is fully installed over the cell holder and measure a blank. Also, check user calibration settings. You might want to check the absorbance reading in 'service mode' and check absorbance values for service test kit vials.
  • C
    Charles AyersSep 2, 2025
    What to do if Hach LPV445.99.09110 Measuring Instruments show error E-00?
    • S
      Shawn RangelSep 2, 2025
      To resolve error E-00 on Hach Measuring Instruments, measure a blank solution to set the instrument to zero. Also, check user settings and reset the mainboard to factory settings.
  • A
    Alicia AndersonSep 4, 2025
    Why does Hach LPV445.99.09110 Measuring Instruments show error E-03?
    • M
      Mitchell CollinsSep 4, 2025
      Error E-03 on Hach Measuring Instruments is due to a configuration or setting error. Prepare a fresh standard. Alternatively, reset the mainboard to factory settings.
  • J
    Jeffrey CarrollSep 6, 2025
    What to do if Hach Measuring Instruments show error E-06?
    • C
      cookdeniseSep 6, 2025
      If your Hach Measuring Instruments display error E-06, it is due to a configuration or setting error. Enter or measure the absorbance value again. You can also try resetting the mainboard to factory settings.

Hach LPV445.99.09110 Specifications

General IconGeneral
ModelLPV445.99.09110
CategoryMeasuring Instruments
ManufacturerHach
Resolution0.01 NTU
Operating Temperature0 to 50°C (32 to 122°F)
Temperature Range0 to 50°C (32 to 122°F)
TypeTurbidity Meter
Measurement Range0 - 1000 NTU
Light SourceLED
MaterialStainless Steel