Questions and Answers

  • C
    Crystal LarsenAug 15, 2025
    What to do if Hach LXV321.99.00001 shows a calibration error?
    • M
      Monica MaxwellAug 15, 2025
      A calibration error on the Hach Measuring Instruments means there is no calibration available, or the calibration points are set incorrectly (upper and lower calibration points transposed). Recalibration is required. Note that a 1-point default calibration of DS curves may not be sufficiently precise for the measuring medium.
  • D
    Daniel RoseAug 18, 2025
    How to fix 'Battery is empty' error on Hach LXV321.99.00001?
    • M
      mayersAug 18, 2025
      If the Hach Measuring Instruments indicates that the battery is empty, it's likely that the battery hasn't been charged. Fully charge the battery to resolve this issue.
  • V
    Vincent RayAug 21, 2025
    Why Hach Measuring Instruments is not communicating with the probe?
    • K
      Kathryn RiceAug 21, 2025
      If your Hach Measuring Instruments is having trouble communicating with the probe, the probe might not be properly connected. Ensure the probe is securely plugged in and that the connection is correct. Verify that the probe number is also correct.
  • A
    Andrew RosalesAug 25, 2025
    What to do if Hach Measuring Instruments displays a probe error?
    • D
      Dustin WrightAug 26, 2025
      If the Hach Measuring Instruments displays a probe error, it indicates a technical issue with the probe. Try switching the instrument off and then back on again. If the message persists, the probe may need to be sent for repair.
  • J
    Jennifer StewartAug 28, 2025
    What to do if Hach LXV321.99.00001 shows an incorrect probe error?
    • A
      Alexander NelsonAug 28, 2025
      If the Hach Measuring Instruments indicates an incorrect probe, it means the probe has been replaced but not re-registered or correctly registered on the instrument. Connect the correct probe and use code 390 to release the DS curve for the new probe, then recalibrate.
  • J
    Justin SmithAug 31, 2025
    What does overflow area mean on Hach LXV321.99.00001?
    • C
      Christina ValdezAug 31, 2025
      If the Hach Measuring Instruments displays an overflow area error, it means the concentration is too high. The probe might be measuring in an unsuitable medium, or it may be necessary to recalibrate.
  • W
    William MacdonaldSep 3, 2025
    What causes offset error on Hach Measuring Instruments?
    • C
      Connie LeeSep 4, 2025
      An offset error on the Hach Measuring Instruments indicates that the calibration point was set incorrectly, specifically too close to the zero point. Recalibration is required to correct this.
  • J
    Jill GreenSep 7, 2025
    How to correct time/date error on Hach LXV321.99.00001?
    • M
      Makayla SteeleSep 7, 2025
      If the Hach Measuring Instruments displays a time/date error, it means the time or date is implausible. Try switching the instrument off and then back on again. Check the time and date settings, and if required, change the clock batteries.
  • C
    Craig BraunSep 12, 2025
    How to fix EEPROM error on Hach LXV321.99.00001?
    • M
      Mr. Michael WilkinsonSep 12, 2025
      To solve EEPROM error on Hach Measuring Instruments, switch the instrument off and then back on again.