EasyManuals Logo

Hach LXV321.99.00001 User Manual

Hach LXV321.99.00001
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
26
Operation
5.1.3 Manual calibration value correction
1. Select MENU, confirm with ENTER/ON.
2. Select
CALIBRATE, confirm with ENTER/ON.
The instrument reads the probe data.
3. Select the curve shown and confirm with ENTER/ON.
4. Select the desired curve and confirm with
ENTER/ON.
5. Select the desired point and confirm with
ENTER/ON.
6. Overwrite the existing calibration point.
7. Enter the value for the calibration point using the navigation keys and confirm with
ENTER/ON.
Navigation key
LEFT/RIGHT: Jump to next/previous decimal place
Navigation key
UP/DOWN: Change number
5.1.4 Delete calibration point
Same procedure as in section 5.1.3. Set the concentration of the calibration point to the
value 0.000 and confirm with
ENTER/ON. The calibration point is deleted.
5.1.5 Reset to default calibration
If the turbidity calibration was changed, it can be reset to the default calibration in
accordance with ISO 7027.
Note: Stop the current measurement before the default calibration is reset.
1. Select MENU, confirm with ENTER/ON.
2. Select
CALIBRATE, confirm with ENTER/ON.
The instrument reads the probe data.
3. Select turbidity curve C-Tu and confirm with ENTER/ON.
Note: If turbidity curve C-Tu is not shown in the menu, select the curve shown and confirm with
ENTER/ON. The selection menu for the curves is shown. Select curve C-Tu and confirm with
ENTER/ON. Curve C-TU is activated.
Open the selection menu for curve C-TU with ENTER/ON, select RESET and confirm with
ENTER/ON.
4. Press navigation key UP/DOWN , select RESET and confirm with ENTER/ON.
The DS curves can also be reset to the default calibration (1-point calibration at 4 g/L).
However, a precise determination of the solids concentration can only be obtained through
calibration (refer to section 5.1, page 23).
5.2 Measure
Note: In order to obtain precise solids measurement results, you must calibrate the instrument
before the first measurement is taken.
Without any calibration of the DS curves, the instrument returns to 1-point default calibration at
4
g/L. Deviations from the actual measurement value may occur.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach LXV321.99.00001 and is the answer not in the manual?

Hach LXV321.99.00001 Specifications

General IconGeneral
BrandHach
ModelLXV321.99.00001
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals