DOC012.97.90371
LZV902
DR 6000
03/2012, Edition 1
User Manual en
Manual del usuario es
Manuel d'utilisation fr
ユーザー マニュアル ja
사용 설명서 ko
Manual do utilizador pt
用户手册 zh

Questions and Answers

  • S
    Sean DonovanSep 14, 2025
    What to do if Hach Measuring Instruments displays 'Web server cannot be reached'?
    • E
      emilywangSep 14, 2025
      If the Hach Measuring Instruments displays 'Web server cannot be reached', it means the Viper homepage cannot be reached. Try the connection again later.
  • J
    Jaime DanielsAug 25, 2025
    What to do if Hach LZV902 Measuring Instruments shows 'Absorbance > 3.5!'?
    • A
      April LongAug 25, 2025
      If the absorbance exceeds 3.5 on your Hach Measuring Instruments, it means the measured absorbance is too high. Dilute the sample and measure it again to get an accurate reading.
  • M
    Michael SchneiderAug 25, 2025
    What to do if Hach LZV902 Measuring Instruments displays 'Concentration too high!'?
    • M
      Mark SmithAug 26, 2025
      If the Hach Measuring Instruments displays 'Concentration too high!', it means the calculated concentration is higher than 999999. Dilute the sample and measure again to resolve this issue.
  • G
    gtaylorAug 26, 2025
    What to do if Hach LZV902 Measuring Instruments says 'Over measuring range'?
    • J
      Joseph OlsonAug 26, 2025
      If your Hach Measuring Instruments indicates that you are over the measuring range, it means the measured absorbance is above the calibration range of the test. To fix this, dilute the sample and measure it again.
  • S
    swoodsSep 14, 2025
    What to do if Hach LZV902 Measuring Instruments says 'Please close the cover'?
    • D
      Daniel PowersSep 15, 2025
      If the Hach Measuring Instruments prompts you to 'Please close the cover', simply close the cover.
  • M
    Michael McdowellSep 23, 2025
    What does 'Control range not reached!' mean on Hach LZV902 Measuring Instruments?
    • R
      robinkingSep 23, 2025
      If the Hach Measuring Instruments displays 'Control range not reached!', this is a warning notice indicating that the control limit set was not reached.
  • D
    Daniel DeleonSep 7, 2025
    What to do if Hach Measuring Instruments says 'Insufficient memory for update'?
    • I
      Isabel CampbellSep 7, 2025
      If the Hach Measuring Instruments displays 'Insufficient memory for update', it means there was an error during the update. Select a memory stick with more space.
  • A
    Austin McmillanSep 8, 2025
    What to do if Hach LZV902 displays 'Error during network drive setup!'?
    • D
      Diana KirkSep 8, 2025
      If the Hach Measuring Instruments displays 'Error during network drive setup!', it indicates an error during network setup because the target directory does not exist. Check the settings and define the target directory.
  • M
    Mr. Joseph MilesSep 23, 2025
    What does 'Control range exceeded!' mean on Hach LZV902?
    • R
      Renee HamiltonSep 23, 2025
      If the Hach Measuring Instruments displays 'Control range exceeded!', this is a warning notice indicating that the control limit was exceeded.
  • M
    Marcus DelgadoAug 31, 2025
    What to do if Hach Measuring Instruments displays 'System check incorrect!'?
    • B
      bporterAug 31, 2025
      If the Hach Measuring Instruments displays 'System check incorrect!', it means the measurement of air values failed. Switch the instrument off and then back on again. If the system check is still unsuccessful, contact technical support.

Hach LZV902 Specifications

General IconGeneral
Accuracy±5% of reading or ±5 mg/L (ppm), whichever is greater
Operating Temperature0 to 60 °C (32 to 140 °F)
Range0 - 4000 mg/L (ppm)
Output4-20 mA, RS-485 Modbus
Measurement Range0 - 4000 mg/L (ppm)
Wetted Materials316 stainless steel