Operation
28
3. Push the MR/DOWN key until Linc displays.
4. Push the CAL/MCLR key. SEL displays with the User icon.
Note: Any additional display of KCl, NaCl or 442 icons indicates a Linc between the
additional solution and the User solution. If none of the solution selection icons are
displayed, nothing is linked to the User mode.
5. Push UP/MS or MR/DOWN to select a standard solution to link to the User
mode calibration constant.
6. Push the CAL/MCLR key to accept the setting. The User mode now uses the
calibration offset constant that was created here.
Note: To exit without changing the setting, push any measurement key.
3.20.3 Cancel User mode calibration Linc
To cancel the User mode calibration Linc:
Note: The MP series meter must be in User linked mode to cancel the “Linc.”
1. Push a (linked) measurement key such as COND, RES, MIN/SALT, or TDS.
Two solutions are displayed on the left side of the display: User and another,
such as KCl.
2. Push the MR/DOWN key until Linc displays.
3. Push the CAL/MCLR key. SEL, User and the linked solution appear on the
display.