EasyManuals Logo

Hach MTC101 User Manual

Hach MTC101
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
Figure 2 Remove the protective tape
Section 5 Calibration
The procedure that follows is applicable to meters that can connect to Intellical ORP (MTC) probes.
Refer to the applicable meter documentation for meter operation and probe-specific settings.
5.1 Calibration notes
Read the notes that follow before calibration:
The meter shows the standard solution to use for calibration. If necessary, change the standard
solution in the probe settings menu.
Do not dilute the standard solution. Discard the standard solution after each calibration.
If the calibration standard is ZoBell's ORP/Redox standard solution, calibrate the probe at a
solution temperature that is between 0 and 30 °C (32 to 86 °F). If the calibration standard is Light’s
solution, calibrate the probe at 25 °C (77 °F).
Use the single display mode for calibration when more than one probe is connected to the meter (if
applicable).
Calibrate the probes and verify the calibration regularly for best results. Use the meter to set
calibration reminders.
The calibration data is stored in the probe. When a calibrated probe is connected to a different
meter with the same calibration options, a new calibration is not necessary.
Air bubbles below the sensor when in solution can cause a slow response or error in the
calibration. Make sure to remove air bubbles during calibration.
Remove the shroud before calibration. Refer to Remove or install the shroud on page 10.
6
English

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach MTC101 and is the answer not in the manual?

Hach MTC101 Specifications

General IconGeneral
BrandHach
ModelMTC101
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals