EasyManua.ls Logo

Hach NT3100sc - 6.1 Maintenance schedule; 6.2 Clean the measuring path

Hach NT3100sc
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
N O T I C E
Do not disassemble the instrument for maintenance. If the internal components must be cleaned or repaired,
contact the manufacturer.
6.1 Maintenance schedule
Table 3 shows the recommended schedule of maintenance tasks. Facility requirements and
operating conditions may increase the frequency of some tasks. Table 4 shows the average service
life of wearing parts for standard operating conditions and default (factory) settings.
Table 3 Maintenance schedule
Task 1 week 3 months 6 months 1 year As necessary
Visual inspection X
Clean the measuring path on page 18 X
Validate the probe calibration on page 21 X
10
Replace the wiper blade on page 19 X
10,11
Manufacturer service inspection X
12
Table 4 Wearing parts consumption
Wearing part Quantity Average service life
Wiper blades (set of 5 pieces) 1 > 1 year
10,11
Wiper motor 1 7 years
11
Wiper shaft with seals 1 2 years
11
Housing seals
13
1 2 years
Flashlamp 1 10 years
Measuring window 2 5 years
10
Filter set 1 5 years
6.2 Clean the measuring path
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
10
Based on non-abrasive water conditions
11
Based on the default settings
12
A minimum of one inspection per year is required. For the best performance and uptime, the
manufacturer recommends 2 inspections per year.
13
Replace each time the probe is opened.
18 English

Table of Contents