Page 9
Calibration Reports
Once a measured gas calibration is completed, a
report is issued showing the log file of the last
actions. The calibration report will contain the
records for the last 10 calibrations. Each
calibration record will contain parameters useful
for traceability. For instance:
the calibration coefficient
all the measurements which influence the
calibration (temperature, barometric
pressure, current...)
the date and time
the user name and ID
Barometric Pressure Calibration
Upper box shows the barometric pressure
measured by the instrument.
Using a precision certified barometer, measure
barometric pressure in the location where the
measuring instrument is used. Compare; if value
are the same press cancel; otherwise enter the
new barometric value in the lower box and
validate the new setting.
Once the calibration is done a calibration report is
created.
Note :
The barometric sensor has been calibrated in
factory.
Calibration (cont)