DOC026.53.80437
SD900 Portable Sampler
02/2015, Edition 2
User Manual

Table of Contents

Other manuals for Hach SD900

Questions and Answers

  • A
    Anna CannonAug 2, 2025
    What to do if the sample volume is not correct in my Hach Recording Equipment?
    • C
      Christopher SingletonAug 2, 2025
      If the sample volume of your Hach Recording Equipment is not correct, several factors could be at play. First, repeat the volume calibration process. Ensure that the correct tube length is specified in the sampler program. Make sure the intake tube is as vertical and short as possible to ensure complete purging. If the strainer is not fully submerged, consider installing a shallow depth strainer. Replace any worn pump tubing or roller assembly. If the liquid sensor is disabled, turn it on and perform a volume calibration. If the liquid sensor isn't operating correctly, calibrate it using the same liquid being sampled.
  • C
    Claudia BrownAug 8, 2025
    How to troubleshoot no instrument power with AC in Hach Recording Equipment?
    • M
      Michael DuarteAug 8, 2025
      If your Hach Recording Equipment has no instrument power with AC, start by verifying AC power at the electrical outlet. If the issue persists, it may be due to a defective controller, and you should contact technical support.
  • L
    Lindsay ChungAug 14, 2025
    Why is the battery life so short on my Hach SD900 Recording Equipment?
    • C
      Christopher BennettAug 14, 2025
      If your Hach Recording Equipment's battery life is short, ensure the lead-acid battery holds between 12.6 V and 13.4 V when fully charged, and replace it if necessary. Fully charge the battery and let it stand for an hour; if the voltage drops below 12.5 V, replace the battery. Also, ensure the correct portable base is selected.
  • R
    Rachel GarciaAug 20, 2025
    What to do if my Hach SD900 Recording Equipment has no instrument power with DC?
    • D
      Debra UnderwoodAug 20, 2025
      If your Hach Recording Equipment has no instrument power with DC, the battery may not be charged or may not hold a charge, so replace the battery or use an AC power cord. If neither of these actions resolves the issue, the controller may be defective, and you should contact technical support.
  • F
    Frank FowlerAug 27, 2025
    How to fix insufficient lift in the Hach SD900 Recording Equipment sampler?
    • L
      Lisa Pena MDAug 27, 2025
      If the sampler on your Hach Recording Equipment does not have sufficient lift, ensure the strainer is completely submerged; if not, install the shallow depth strainer. Check for leaks in the intake tube and replace it if necessary. Also, examine the pump tube for wear and replace it if needed. If the pump roller assembly is worn, contact technical support.

Summary

Specifications

General Information

Overview of the sampler's general specifications and safety information.

SD Controller Specifications

Technical specifications for the SD controller unit, including dimensions, weight, and power.

Installation

Confined Space Precautions

Safety guidelines and definitions related to installing the sampler in confined spaces.

Mechanical Installation

Guidelines for physical setup, including site selection and mounting the sampler.

Electrical Installation

Instructions for connecting power and other electrical components to the sampler.

User Interface and Operation

Controller Keypad

Description of the buttons and layout of the controller keypad for operation.

Main Menu Overview

Explanation of the main menu options for sampler operation and data management.

Configure System Settings

How to adjust system parameters like date, time, language, and communication settings.

Configure the SDI-12 Device

Steps to set up and manage connected SDI-12 sensors and their measurements.

Set Up a Sampling Program

Guide to creating, modifying, and saving sampling programs with various parameters.

Manual Operation

Procedures for manual control of the sampler, including grab samples and pump operation.

View Data

How to access and interpret sample history, event logs, and SDI-12 device data.

Volume Calibration

Procedures for calibrating sample volume, either automatically with a liquid sensor or manually.

Maintenance

Clean the Instrument

Instructions for cleaning the sampler's enclosure and components.

Replace the Desiccant

Guide on how to replace the desiccant cartridge to prevent corrosion.

Pump Maintenance

Procedures for maintaining the pump, including replacing the pump tubing.

Troubleshooting

General Troubleshooting

Table listing common problems, their causes, and solutions for the sampler.

Error Messages

Explanation of error messages displayed by the controller and how to resolve them.

Replacement Parts and Accessories

Bottle Kits

List of available bottle kits for different sample volumes and configurations.

Containers and Bottle Sets

Catalog of various containers and pre-configured bottle sets.

Replacement Parts

List of individual replacement parts for the sampler.

Accessories

Catalog of optional accessories and cables for the sampler.

Hach SD900 Specifications

General IconGeneral
BrandHach
ModelSD900
CategoryRecording Equipment
LanguageEnglish