EasyManua.ls Logo

Hach SD900 - User Interface and Operation

Hach SD900
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Connect an SDI-12 device
N O T I C E
Measurement errors can occur due to electrical transients from sources such as lightning or large electrical
motors. These errors can cause unexpected sample acquisition or a missed sample when the controller is
programmed for setpoint sampling based on sonde measurements. Use a power line filter or connect the
controller to a different branch circuit to reduce the possibility of transients.
The controller supplies power to the SDI-12 device. Use the 15 m (50 ft) cable or the 30 m (100 ft)
cable to connect a sonde to the controller. Refer to Figure 10.
Note: A separate power supply for the sonde is necessary for data logging by the sonde.
The SDI-12 interface is enabled only when a connected device is found by the controller. The scan
for a sensor occurs automatically. The address that is assigned to the SDI-12 device is automatically
detected by the controller.
Figure 10 Sonde connection
1 Sonde 4 SDI-12 connector
2 Sonde bulkhead connector 5 15 or 30 m (50 or 100 ft) cable
3 6-pin cable connector
Connect the sampler to power
D A N G E R
Electrocution hazard. If this equipment is used outdoors or in potentially wet locations, a Ground Fault
Circuit Interrupt (GFCI/GFI) device must be used for connecting the equipment to its main power
source.
D A N G E R
Fire hazard. Install a 15 A circuit breaker in the power line. A circuit breaker can be the local power
disconnect, if located in close proximity to the equipment.
D A N G E R
Electrocution hazard. Protective Earth Ground (PE) connection is required.
W A R N I N G
Electrocution hazard. Make sure that there is easy access to the local power disconnect.
14 English

Table of Contents

Other manuals for Hach SD900