12
6. Operation
6.3.3. Two- and three-point calibration
Follow the diagram below.
The equipment will repeat the process described for the first stan-
dard solution with the second and third.
Dip the sensor into the tube several times to homogenize the li-
quid. Do not hold it by the tube to avoid heating of the standard
solution.
Note: if the measuring instrument is not used for 5 minutes, it shuts
off automatically.
If errors occur during calibration, an error message will appear
on screen (see pg. 19).
When you switch standards, rinse the cell with deionized water.
μS/cm
μS/cm
ºC
T
t
μS/cm
EC
ºC
T
t
μS/cm
EC
μS/cm
ºC
T
t
μS/cm
EC
ºC
T
t
μS/cm
EC
@25ºC
μS/cm
ºC
EC
T
t
50 60
0.2... 200.000 μS/cm -30... 80ºC
50 60
0.2... 200.000 μS/cm -30... 80ºC
50 60
0.2... 200.000 μS/cm -30... 80ºC
50 60
50 60
0.2... 200.000 μS/cm -30... 80ºC
50 60
Screw the
sensor
to the 1st tube
Screw the
sensor to the
2nd tube
To exit calibration
Two-point calibration
To exit calibration.
One-point calibration
sensION+ EC5 HACH 0710.indd ANG:12sensION+ EC5 HACH 0710.indd ANG:12 7/9/10 12:51:007/9/10 12:51:00