2-24 System Description
P/N 39551-00, Rev. F
Warning! The bowl base must be firmly installed and evenly
seated in the centrifuge chuck. If the centrifuge chuck spins with
the bowl base not evenly seated, as indicated by bowl wobbling
or noise, bowl damage will occur and the procedure must be
discontinued.
Do not attempt to reseat the bowl or continue the procedure.
Do not return the contents of the bowl to the donor. Contact
Haemonetics Technical Hot Line at (800) 537-2802, and dis
-
card the bowl contents and disposable per the facility’s SOP.
Figure 2-11, Mechanical Centrifuge Components
Cover
Bowl Optics
Fluid Detectors
Mechanical Chuck
Mechanical Clips
Cover Latch
Centrifuge Well