Notice Messages and Troubleshooting 5-7
P/N 39551-00, Rev. F
In case the procedure is to be discontinued:
10. Once all blood has been returned and the MCS+ device initiates the next
Draw mode, press the STOP key to discontinue the procedure. Press the
HELP key to enter the HaemoUpdate screen and obtain procedure statistics.
In case the procedure is to be continued:
11. Allow the MCS+ device to continue the procedure per standard procedures.
GRAVITY REINFUSION PROCEDURE
In the rare event that it is not possible to automatically return the blood in the
disposable (e.g., power outage), and it is deemed necessary to return the red cells
from the bowl, use the following procedure. The collection targets (Number of
platelets, volume of plasma, etc.) for the procedure may not have been met. Refer
to the HaemoUpdate screen for current product information. In case of a general
power loss, this information may be retrieved by performing a procedure recov
-
ery when power has been restored. The disposable does not need to remain on
the machine. To retrieve the information, press HELP at the “Install Disposable”
screen for procedure statistics.
Note: To maintain 5-day storage of platelet products, the platelet product must
be hermetically sealed and removed from the disposable prior to removing the
venipuncture needle. Refer to the appropriate procedures given in the Protocol
Manual for sealing and removing the platelet bags.
1. Hemostat the donor line near the needle.
2. Remove white-striped tubing from white valve, or applicable line to bag con-
taining the air from the bowl.
3. Remove the product bags per the procedures given in the appropriate section
of the applicable protocol manual (platelet and plasma procedures only).
4. Remove the red-striped tubing from the Red Valve.
5. Remove white cartridge and Blood and White Pump tubing.
6. Release the hemostat near needle and observe flow from bowl to needle. Roll
air bag to assist flow. If flow slows or stops, remove bowl from centrifuge and
hold upright higher than donor’s heart to facilitate flow.
7. Continue to infuse until all blood is returned and then clamp the donor line
and remove needle.
8. Press the HELP key to enter the HaemoUpdate screen to obtain applicable statistics.
9. Record per local SOP.

Table of Contents

Other manuals for Haemonetics MCS+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Haemonetics MCS+ and is the answer not in the manual?

Haemonetics MCS+ Specifications

General IconGeneral
BrandHaemonetics
ModelMCS+
CategoryMedical Equipment
LanguageEnglish

Related product manuals