5-8 Notice Messages and Troubleshooting
P/N 39551-00, Rev. F
Table 5-1, Notice Messages
*
# Notice Messages Help Messages
0
SYSTEM PROBLEM
THE COMPUTER HAS ENCOUNTERED AN ERROR.
A CPU error caused the MCS+ device to reset.
1. Power off and complete a
Procedure Recovery.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
5
BLAD DETECTS AIR
BLOOD LINE AIR DETECTOR
BLOOD LINE AIR DETECTOR
1. Check tubing installation in BLAD.
2. Check donor line for flow restriction.
3. Resume operation after air is cleared from BLAD.
6
DLAD1 DETECTS AIR
DONOR LINE AIR DETECTOR 1
DONOR LINE AIR DETECTOR 1
1. Check tubing installation in DLAD1.
2. If air has traveled past DLAD1 or DLAD2, open
the RED valve and turn the BLOOD pump
CLOCKWISE to move air into the bowl.
3. Resume operation after air is cleared from tubing
back to bowl.
7
DLAD2 DETECTS AIR
DONOR LINE AIR DETECTOR 2
DONOR LINE AIR DETECTOR 2
1. Check tubing installation in DLAD2.
2. If air has traveled past DLAD1 or DLAD2, open
the RED valve and turn the BLOOD pump
CLOCKWISE to move air into the bowl.
3. Resume operation after air is cleared from tubing
back to bowl.
15
thru
24
SYSTEM PROBLEM
THE COMPUTER HAS ENCOUNTERED AN ERROR.
A CPU error caused the MCS+ device
to reset.
1. Power off and complete a
Procedure Recovery.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
25
CENTRIFUGE PROBLEM
THE CENTRIFUGE IS NOT TURNING.
1. Check that bowl is installed properly.
2. Check that Centrifuge rotates freely.
3. Check header of bowl.
4. Resume operation.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
26
CENTRIFUGE PROBLEM
THE CENTRIFUGE WAS TURNING TOO FAST.
1. Wait until Centrifuge has stopped.
2. Check Centrifuge for obstructions.
3. Resume operation.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
*
Platelet protocols in Platelet Protocol Software Version E or later have revised notice messages and associated message
numbers. Refer to Appendix C of the Platelet Protocol Manual, Part Number 51052-00 for details.

Table of Contents

Other manuals for Haemonetics MCS+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Haemonetics MCS+ and is the answer not in the manual?

Haemonetics MCS+ Specifications

General IconGeneral
BrandHaemonetics
ModelMCS+
CategoryMedical Equipment
LanguageEnglish

Related product manuals