Notice Messages and Troubleshooting 5-9
P/N 39551-00, Rev. F
27
CENTRIFUGE PROBLEM
THE CENTRIFUGE VACUUM IS NOT DETECTED.
The vacuum is too low to hold the bowl.
1. Remove bowl from the Centrifuge.
2. Re-seat bowl.
3. Ensure that bowl is properly seated.
4. Resume operation.
28
SYSTEM PROBLEM
FLUID HAS BEEN DETECTED IN THE CENTRIFUGE.
A leak may have occurred.
This procedure MUST be discontinued!
1. Discontinue the procedure.
2. Note the condition of the disposable
before disposing of it.
3. Contact Haemonetics Technical Hot Line.
NOTE: The disposable should be returned to
Haemonetics for analysis.
29
30
SYSTEM PROBLEM
THE COMPUTER HAS ENCOUNTERED AN ERROR.
A CPU error caused the MCS+ device to reset.
1. Power off and complete a
Procedure Recovery.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
31
CENTRIFUGE PROBLEM
THE CENTRIFUGE SPEED WAS TOO SLOW.
1. Wait until Centrifuge has stopped.
2. Check that bowl is installed properly.
3. Check Centrifuge for obstructions.
4. Resume operation.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
32
thru
36
SYSTEM PROBLEM
THE COMPUTER HAS ENCOUNTERED AN ERROR.
A CPU error caused the MCS+ device to reset.
1. Power off and complete a
Procedure Recovery.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
37
AIR DETECTED BY
AIR DETECTED BY SYSTEM MONITOR.
The Air Detector detects air.
1. Check tubing installation in detector.
2. Turn AC or Blood Pump CLOCKWISE
to remove air.
3. Resume operation.
38
SYSTEM PROBLEM
THE COMPUTER HAS ENCOUNTERED AN ERROR.
A CPU error caused the MCS+ device to reset.
1. Power off and complete a
Procedure Recovery.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
39
CENTRIFUGE PROBLEM
THE CENTRIFUGE SPEED WAS TOO SLOW.
1. Wait until Centrifuge has stopped.
2. Check that bowl is installed properly.
3. Check Centrifuge for obstructions.
4. Resume operation.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
# Notice Messages Help Messages

Table of Contents

Other manuals for Haemonetics MCS+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Haemonetics MCS+ and is the answer not in the manual?

Haemonetics MCS+ Specifications

General IconGeneral
BrandHaemonetics
ModelMCS+
CategoryMedical Equipment
LanguageEnglish

Related product manuals