5-10 Notice Messages and Troubleshooting
P/N 39551-00, Rev. F
40
PUMP PROBLEM
AC PUMP WAS TURNING TOO SLOW.
A restriction may have occurred.
1. Check pump for restrictions.
2. Check pump for kinked tubing.
3. Resume operation.
41
PUMP PROBLEM
BLOOD PUMP WAS TURNING TOO SLOW.
A restriction may have occurred.
1. Check pump for restrictions.
2. Check pump for kinked tubing.
3. Resume operation.
42
PUMP PROBLEM
WHITE PUMP WAS TURNING TOO SLOW.
A restriction may have occurred.
1. Check pump for restrictions.
2. Check pump for kinked tubing.
3. Resume operation.
44
PUMP PROBLEM
AC PUMP WAS TURNING TOO FAST.
A restriction may have occurred.
1. Check pump for restrictions.
2. Check pump for kinked tubing.
3. Resume operation.
45
PUMP PROBLEM
BLOOD PUMP WAS TURNING TOO FAST.
A restriction may have occurred.
1. Check pump for restrictions.
2. Check pump for kinked tubing.
3. Resume operation.
46
PUMP PROBLEM
WHITE PUMP WAS TURNING TOO FAST.
A restriction may have occurred.
1. Check pump for restrictions.
2. Check pump for kinked tubing.
3. Resume operation.
48
PUMP PROBLEM
AC PUMP WAS TURNING IN THE WRONG DIRECTION.
A restriction may have occurred.
1. Check pump for restrictions.
2. Check pump for kinked tubing.
3. Resume operation.
49
PUMP PROBLEM
BLOOD PUMP WAS TURNING IN THE WRONG DIRECTION.
A restriction may have occurred.
1. Check pump for restrictions.
2. Check pump for kinked tubing.
3. Resume operation.
50
PUMP PROBLEM
WHITE PUMP WAS TURNING IN THE WRONG DIRECTION
A restriction may have occurred.
1. Check pump for restrictions.
2. Check pump for kinked tubing.
3. Resume operation.
52
PUMP PROBLEM
AC PUMP POSITION IS NOT CORRECT.
A restriction may have occurred.
1. Check pump for restrictions.
2. Check pump for kinked tubing.
3. Resume operation.
53
PUMP PROBLEM
BLOOD PUMP POSITION IS NOT CORRECT
A restriction may have occurred.
1. Check pump for restrictions.
2. Check pump for kinked tubing.
3. Resume operation.