5-4 Notice Messages and Troubleshooting
P/N 39551-00, Rev. F
If an error condition occurs, the operator should record the message number for
later reference. The operator does not have to record the notice text (in the above
example, this is “ACAD Detects Air ”). Pressing NO mutes the alarm for a period
of up to one minute, and pressing HELP displays a Help screen with further trou
-
bleshooting hints. A typical Help screen is shown below.
Automatic monitoring of the MCS+ device begins the moment the system is pow-
ered on. The first action of the MCS+ device is to test its own components and
ensure that the correct disposable set has been installed properly.
Failure of the self-test halts the test sequence. Should this occur, record the test
failure message, power off the machine, and power on. If the problem persists,
contact a Haemonetics Field Service Engineer.
SDPS
Filling Bowl
Press NO to Mute the Alarm.
ANTICOAGULANT AIR DETECTOR
Press HELP for the HaemoUpdate display.
Hints
1. Check tubing installation in ACAD.
2. If air is present, turn AC pump CLOCKWISE to
4. Resume operation after air is cleared from ACAD.
Help
Messag
e
DRAW
175
Message
Number
Description
of Problem
move air past ACAD.
3. Call Haemonetics Technical Service if problem persists.

Table of Contents

Other manuals for Haemonetics MCS+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Haemonetics MCS+ and is the answer not in the manual?

Haemonetics MCS+ Specifications

General IconGeneral
BrandHaemonetics
ModelMCS+
CategoryMedical Equipment
LanguageEnglish

Related product manuals