Notice Messages and Troubleshooting 5-17
P/N 39551-00, Rev. F
372
SYSTEM PROBLEM
THE PROTOCOL CARD DOES NOT MATCH DISPOSABLE.
The installed Protocol Card and disposable
do not match.
1. Check that the correct disposable
is installed.
2. Check that the correct Protocol Card
is inserted.
3. Clean the disposable ID window.
373
PRESSURE PROBLEM
NO DONOR PRESSURE CHANGE DETECTED AT THE DPM.
The DPM cannot detect venous pressure
change.
1. Check the DPM tubing for occlusions
2. Check the DPM tubing for clamps
3. Ensure the DPM filter is dry.
4. Resume operation.
374
PRESSURE PROBLEM
NO SYSTEM PRESSURE CHANGE DETECTED AT THE SPM.
1. Check the SPM tubing for occlusions.
2. Check the SPM tubing for clamps.
3. Ensure the SPM filter is dry.
4. Resume operation.
NOTE: The product shelf life becomes
24 hrs if the SPM filter is wet.
375
ACAD AIR REMOVED
ANTICOAGULANT AIR DETECTOR
ANTICOAGULANT AIR DETECTOR
1. Check tubing installation in ACAD.
2. If air is present, turn AC pump CLOCKWISE to
move air past ACAD.
3. Call Haemonetics Technical Service if problem
persists.
4. Resume operation after air is cleared from ACAD.
376
BLAD AIR REMOVED
BLOOD LINE AIR DETECTOR
BLOOD LINE AIR DETECTOR
1. Check tubing installation in BLAD.
2. Check donor line for flow restriction.
3. Open the RED valve and turn the BLOOD pump
CLOCKWISE to move air into the bowl.
4. Resume operation after air is cleared from BLAD.
377
DLAD1 AIR REMOVED
DONOR LINE AIR DETECTOR 1
DONOR LINE AIR DETECTOR 1
1. Check tubing installation in DLAD1.
2. If air has traveled past DLAD1 or DLAD2, open
the RED valve and turn the BLOOD pump
CLOCKWISE to move air into the bowl.
3. Resume operation after air is cleared from tubing
back to bowl.
378
DLAD2 AIR REMOVED
DONOR LINE AIR DETECTOR 2
DONOR LINE AIR DETECTOR 2
1. Check tubing installation in DLAD2.
2. If air has traveled past DLAD1 or DLAD2, open
the RED valve and turn the BLOOD pump
CLOCKWISE to move air into the bowl.
3. Resume operation after air is cleared from tubing
back to bowl.