Notice Messages and Troubleshooting 5-21
P/N 39551-00, Rev. F
419
VALVE PROBLEM
GREEN VALVE SELF TEST FAILED.
1. Toggle the valve manually.
2. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
420
SELF TEST PROBLEM
A/D CONVERTER 20% TEST FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
421
SELF TEST PROBLEM
A/D CONVERTER 50% TEST FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
422
SELF TEST PROBLEM
A/D CONVERTER 80% TEST FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
423
SELF TEST PROBLEM
PLEASE CLAMP THE DPM LINE AND REMOVE FILTER.
During self test, the DPM and SPM filters
can not be installed.
1. Clamp tubing to the appropriate filter.
2. Remove filter from connector.
3. Resume operation.
424
SELF TEST PROBLEM
PLEASE CLAMP THE SPM LINE AND REMOVE FILTER.
During self test, the DPM and SPM filters
can not be installed.
1. Clamp tubing to the appropriate filter.
2. Remove filter from connector.
3. Resume operation.
425
SELF TEST PROBLEM
Please install the DPM Filter.
Please install the filter.
1. Attach filter.
2. Ensure that the filter is not wet.
3. Resume operation.
426
SELF TEST PROBLEM
Please install the SPM Filter.
Please install the filter.
1. Attach filter.
2. Ensure that the filter is not wet.
3. Resume operation.
427
SELF TEST PROBLEM
THE DPM CONNECTOR TEST FAILED.
The MCS+ device could not detect the pressure
monitor filter was removed.
1. Remove pressure filter from connector.
2. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
428
SELF TEST PROBLEM
THE SPM CONNECTOR TEST FAILED.
The MCS+ device could not detect the pressure
monitor filter was removed.
1. Remove pressure filter from connector.
2. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.