SPECIFICATIONS
4. SPECIFICATIONS
4.1. COMPARISON TABLE FOR BL12X SWIMMING POOL CONTROLLERS
pH measurement
ORP measurement
Acid dosing pump
Chlorine dosing pump
Analog outputs
Hanna Cloud connectivity
BL120
BL121
BL122
BL123
4.2. TECHNICAL SPECIFICATIONS OF BL120, BL121 & BL122, BL123
Range
0.00 to 14.00 pH
±2000 mV
-5.0 to 105.0 ºC (23.0 to 221.0 °F)*
Resolution
0.01 pH
1 mV
0.1 °C (0.1 ºF)
Accuracy
@25 °C / 77 °F
±0.05 pH
±5 mV
±1.0 ºC (±1.8 ºF)
Calibration
pH buffer calibration: automatic, two points (4.01, 7.01, 10.01 pH)
pH process calibration: adjustable, single point
ORP (mV) calibration: adjustable, single point
Temperature compensation Automatic -5.0 to 105.0 ºC (23.0 to 221.0 ºF) for pH
pH Controller
Proportional feed using adjustable set point and adjustable proportional band
Delay to start at power-on
Overdosing protection using overfeed safety timer
ORP Controller
Proportional feed using adjustable set point and adjustable proportional band
Delay to start at power-on
Overdosing protection using overfeed safety timer
pH regulator interlocked
Alarms
High and Low with enable / disable option for all parameters
Alarm is triggered after five consecutive readings over / under threshold
*The range (pH & temperature) may be limited by the probe’s limits.