CALIBRATION
If the buffer is not recognized (either because the pH electrode has not been placed in solution or the
reading is outside accepted range), “---- WRONG” message is displayed along with CAL tag blinking.
After pH 4.01 or 10.01 buffer is accepted, the “SAVE” message is displayed and the controller
returns to measurement mode.
The “CAL” tag is displayed automatically in measurement mode after a calibration has been performed.
Note: To perform one-point calibration using pH 7.01 buffer, press CAL key after the buffer is saved.
Two-Point Calibration
• Follow One-Point Calibration section for one-point calibration using pH 7.01.
• After pH 7.01 has been accepted, the “pH 4.01 USE” message is displayed.
Place the electrode in the second calibration buffer (pH 4.01 or 10.01); it will be automatically
recognized. After the second buffer has been accepted, the “SAVE” message is displayed for
1 second and the controller returns to measurement mode.