CALIBRATION
10. CALIBRATION
10.1. pH CALIBRATION
The HI981412 provides a digital calibration at the push of a button. Calibrate the probe frequently
for improved accuracy. Also:
• before in-line or flow cell installation
• whenever the probe is replaced
• after periodic maintenance
Always use fresh calibration buffers and perform electrode maintenance prior to calibration
(see ELECTRODE CARE & MAINTENANCE section) .
Preparation
Pour the buffer solution into clean beakers. If possible, use plastic beakers to minimize any
EMC interferences. For accurate calibration and to minimize cross-contamination, use two beakers,
one for rinsing the probe and one for calibration.
User Calibration
One- or two-point calibration can be performed using one of the three standard buffer solutions: 4.01,
7.01 or 10.01pH.
When a two-point calibration is required, use 4.01 pH buffer as first calibration point.
Procedure
• Press CAL key to enter calibration mode.
• Place the electrode in chosen pH buffer solution.
• Press CAL key to save calibration point and return to measurement screen.
• Long press MENU key to exit without saving.
One-Point Calibration
Press CAL key to enter calibration mode. “pH 7.01 USE” message is displayed as calibration point.
Place the electrode in 4.01, 7.01 or10.01 pH buffer. The controller automatically recognizes the
buffer value.
When the buffer is recognized, “RECOGNIZED” is scrolled at the bottom of the LCD display. “WAIT
is displayed until the reading is stable and the calibration is accepted.