MAINTENANCE
13. MAINTENANCE
13.1. ELECTRODE CARE & MAINTENANCE
Proper care and maintenance of the pH probe is essential for accurate readings. Cleaning, calibrating,
and appropriate storage will extend the life of the probe.
Remove the electrode protective cap. Do not be alarmed if any salt deposits are present, this is
normal. Rinse the probe with water.
Shake the electrode down as you would do with a clinical thermometer to eliminate any air bubbles
inside the glass bulb (pH electrode only).
If the bulb and/or junction are dry, soak the electrode in HI70300 storage solution for a minimum
of 30 minutes. Rinse with water.
• Calibrate before using.
When the electrode is not in use, add a few drops of HI70300 storage solution to the protective
cap and replace the cap. To ensure a quick response, the glass bulb (pH electrode) and the junction
should be kept moist and not allowed to dry out. This can be achieved by installing the electrode in
such a way that it is constantly in the flow cell or the pipe filled with the sample.
Note: Never store the electrode in distilled or deionized water.
13.2. PUMP TUBING REPLACEMENT
For pump tubing replacement please wear protective gloves and eye protection at all times; and follow
the steps below:
1. Power off the controller.
2. Disconnect the tubing from the pump.
3. Remove the plastic cover from the pump, which is fixed with a screw.