SETUP
7. SETUP
• Short press MENU key to move to next item in menu.
• Long press MENU key to exit.
• Press arrow keys to change the values.
• Short press MENU key to automatically save modified values.
Table below presents an overview of the menu with ranges and factory set defaults.
Parameter
(Scrolled message)
Range / Option Default settings
Control Auto or oFF Auto
Control type on/oFF or Proportional on/oFF
Control mode Hi or Lo Hi
Set point 4.00 to 10.00 7.20
Hysteresis (on/oFF only) 0.10 to 1.00 pH 0.50
Band (Proportional only) 0.10 to 2.00 pH 1.0
Start up delay (seconds) 0-600 60
Overtime alarm (minutes) oFF, 0-180 30
Flow rate (Liter/hour)
Flow rate (Gallon/hour)
0.5 to 3.5
0.13 to 0.92
1.0
0.26
Level alarm diS or En diS
High alarm diS or En En
High alarm value (pH) 0 to 14.00 pH* 10.00 pH
Low alarm diS or En diS
Low alarm value (pH) 0 to 13.90 pH* 4.00 pH
Temperature unit °C or °F °C
Flow rate unit L.H or GAL.H L.H
Language En (English), ES (Spanish),
Fr (French), Pt (Portuguese),
nL (Dutch), dE (German)
En (English)
*Available range changes based upon other settings.
High alarm must be set higher than low alarm. If low alarm was set 7 pH, then high alarm range is
7.1 to 14 pH.