10
If LOCK was pressed to freeze the LCD range and the reading went out of
range, the full-scale value of the frozen range will be displayed blinking.
The EC reading is affected by temperature.
Three options of compensating temperature are available, on EC measurement
mode.
Note: The compensation is referenced at the selected reference temperature
(see SETUP for details, page 31).
Automatic (Atc): The EC probe has a built-in temperature sensor; the
temperature value is used to automatically compensate the EC/TDS reading
(from 0.0  60.0 ºC).
Manual (Mtc): The temperature value, shown on the secondary LCD, can be
manually set with the ARROW keys. The ºC tag blinks when this option is active.
No Compensation (notc): The temperature
value is displayed, but is not taken into account
and ºC tag blinks with a higher frequency.
The reading displayed on the primary LCD is the
actual EC or TDS value. To select the desired
option, press the ALT&ATC keys until the option
is displayed on the LCD.
Note: The default compensation mode is Atc.
If no temperature probe detected, Atc mode can not be selected
and the instrument displays ---- on the secondary LCD.
If temperature compensation is active, measurements are compensated using the
temperature coefficient (default value 1.90 %/ºC). To change the temperature
coefficient, enter the setup mode and select the tc item (see SETUP for details,
page 31). The current temperature coefficient can be quickly viewed by pressing
the ALT&TC keys. The value is briefly displayed on the secondary LCD.
If the temperature reading is out of 0.0  60.0 ºC interval and Atc option is
selected, the ºC tag will blink and the closest interval limit will be displayed.
Press the ARROW keys to change the displayed temperature value. This
value is used to compensate the EC/TDS reading.
TDS MEASUREMENTS
Press the ALT&MODE keys while in EC range. The instrument will switch to TDS
measuring range. A TDS reading will be displayed on the primary LCD and the
temperature reading on the secondary LCD.