6564
electrode condition, with sign (3 chars). The
“-01” code means not calculated
ISE calibration data (if available) which contains:
the number of calibrated standards (1 char)
the ISE unit (2 chars)
the calibration slope, with sign and decimal point
(7 chars)
the calibration time of the standard, yymmdd
hhmmss (12 chars)
standards information (for each standard)
type (1 char): 0 - always standard solution
status (1 char): N (new) - calibrated in last
calibration; O (old) - from an old calibration
warnings during calibration (2 chars): 00 -
no warning
standard value, with sign and decimal point
and exponent (12 chars)
calibration time, yymmddhhmmss (12 chars)
Rel mV calibration data (if available), which contains:
the calibration offset, with sign (7 chars)
the calibration time, yymmddhhmmss (12 chars)
Channel 2 GLP
EC calibration data (if available), which contains:
the number of calibrated standards (1 char)
the offset, with sign and decimal point (7 chars)
the cell constant, with sign and decimal point (7 chars)
the calibration time, yymmddhhmmss (12 chars)
standards information (for each buffer)
the unit of the standard (0 - microsiemens,
1 - milisiemens)
the standard value, with sign, decimal
point and exponont (11 chars)
the calibration time of the standard,
yymmddhhmmss (12 chars)
the temperature reference value (2 chars)
the temperature compensation mode (2 chars:
00 - NoTC, 01 - MTC, 02 - ATC)
the temperature compensation coefficient, with
The unit of the second reading (1 char) (it is sent
only for Resistivity, TDS and Salinity ranges only). It
represents the unit of the EC reading (0 - microsiemens,
1 - milisiemens)
Secondary reading (only when primary reading is
not mV or EC) - 8 ASCII chars, including sign and
decimal point.
Temperature reading - 7 ASCII chars, with sign and
decimal point, always in ºC.
MDR Requests the instrument model name and firmware code
(16 ASCII chars).
GLPx Requests the calibration data record.
x=1 calibration data from channel 1
x=2 calibration data from channel 2
The answer string contains:
GLP status (1 char): represents a 4 bit hexadecimal
encoding.
0x01 - pH calibration available (Ch1) or EC
calibration available (Ch2)
0x02 - Rel mV calibration available (Ch1) or
Salinity calibration available (Ch2)
0x04 - ISE calibration available
Channel 1 GLP
pH calibration data (if available), which contains:
the number of calibrated buffers (1 char)
the offset, with sign and decimal point (7 chars)
the average of slopes, with sign and decimal
point (7 chars)
the calibration time, yymmddhhmmss (12 chars)
buffers information (for each buffer)
type (1 char): 0 - standard, 1 - custom
status (1 char): N (new) - calibrated in last
calibration; O (old) - from an old calibration.
warnings during calibration (2 chars): 00 -
no warning, 04 - Clean Electrode warning
buffer value, with sign and decimal point and
exponent (12 chars)
the calibration time of the buffer, yymmdd
hhmmss (12 chars).