HOW TO ENSURE ACCURATE MEASUREMENTS
Each time the cuvette is used, tighten the cap to the same
degree.
Discard the sample soon after the reading is taken to avoid
permanently clouding the glass.
All glassware used to contain the standards and the samples
should be maintained clean, washed with HI 93703-50
cleaning solution and rinsed with HI 93703-0 or turbidity-free
water.
Collect the samples in clean glass or plastic bottles, fit stoppers
and perform the analysis quickly. If unavoidable, store the
sample in a cool, dark place, but not for longer than 24 hours
(the sample needs to be kept at room temperature prior to the
analysis).
To obtain a representative sample, gently, but thoroughly, mix
it before samples are taken. Do not shake (to prevent air
bubbles) and do not let the sample settle.
It is recommended to monthly calibrate the meter with the
supplied HI 93703-10 @10 FTU* standard or more frequently
for greatest accuracy.
Before inserting vials into the
instrument, wipe them with
HI 93703-70 soft, lint-free
tissue. Handle vials so that no
fingerprints can get on the
areas where light passes
(approximately 2 cm/1 inch
from the bottom of the vial).
If you experience any problems in taking measurements, please
contact your dealer or the nearest Hanna Instruments Customer
Service Department.
* 1 FTU = 1 NTU