EasyManuals Logo

Hanna Instruments HI 93703 User Manual

Hanna Instruments HI 93703
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #19 background imageLoading...
Page #19 background image
19
All components have been selected to minimize current drain
without compromising functionality.
In order to minimize the battery consumption, the meter is
equipped with an auto-off function which switches the meter off
after 5 minutes of non-use.
The power source are 4 alkaline batteries (1.5V AA) with an
expected life of 60 hours, or 900 measurements.
To always grant accurate measurements, the batteries are monitored
to ensure that readings are not taken when there is insufficient
power.
A “LO BAT“ indication will appear on the lower right corner of the
display when the batteries are weak and require replacement. The
instrument can still perform approx. 50 measurements.
A “-BA-“ indication will appear on the display when the batteries
are too weak to perform reliable measurements. The message
appears for a few seconds, and then the meter will automatically
switch off. Batteries must be immediately replaced.
BATTERY REPLACEMENT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 93703 and is the answer not in the manual?

Hanna Instruments HI 93703 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 93703
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals