14
1.6 SPECIFICATIONS
TEMPERATURE
RangeRange
RangeRange
Range -5.00 to 55.00 °C;
23.00 to 131.00 °F; 268.15 to 328.15 K
ResolutionResolution
ResolutionResolution
Resolution 0.01 °C; 0.01 °F; 0.01 K
AccuracyAccuracy
AccuracyAccuracy
Accuracy ± 0.15 °C; ± 0.27 °F; ±0.15 K
CalibrationCalibration
CalibrationCalibration
Calibration Automatic at 1 custom point
pHpH
pHpH
pH
RangeRange
RangeRange
Range 0.00 to 14.00 pH; ± 600.0 mV
ResolutionResolution
ResolutionResolution
Resolution 0.01 pH; 0.1 mV
AccuracyAccuracy
AccuracyAccuracy
Accuracy ± 0.02 pH; ± 0.5 mV
CalibrationCalibration
CalibrationCalibration
Calibration Automatic 1, 2 or 3 points with 5 memorized
standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01)
or 1 custom buffer
ORP
RangeRange
RangeRange
Range ± 2000.0 mV
ResolutionResolution
ResolutionResolution
Resolution 0.1 mV
AccuracyAccuracy
AccuracyAccuracy
Accuracy ± 1.0 mV
CalibrationCalibration
CalibrationCalibration
Calibration Automatic at 1 custom point
DISSOLVED OXYGEN
RangeRange
RangeRange
Range 0.0 to 500.0 %
0.00 to 50.00 mg/L
ResolutionResolution
ResolutionResolution
Resolution 0.1 %
0.01 mg/L
AccuracyAccuracy
AccuracyAccuracy
Accuracy 0.0 to 300.0 %: ± 1.5 % of reading
or ± 1.0% whichever is greater;
300.0 to 500.0 %: ± 3% of reading
0.00 to 30.00 mg/L: ± 1.5 % of reading
or 0.10 mg/L whichever is greater;
30.00 mg/L to 50.00 mg/L: ± 3% of reading
CalibrationCalibration
CalibrationCalibration
Calibration Automatic 1 or 2 points at 0, 100 % or 1 custom point