Restore factory settings
This item allows to reset all “System Setup” and
“Measurement Setup” parameters to their de-
fault values.
• Select the “Restore factory settings” item and
press <OK>.
• The meter will ask to confirm: press <Yes> to confirm or <No> to escape.
NoteNote
NoteNote
Note To quit the “System Setup” mode at any time, press ESC. For all items
for which the new selections have been not confirmed, the meter will
keep the previous settings.
3.3 TABLE OF MEASUREMENT AND SYSTEM SETUP ITEMS
Measurement SetupMeasurement Setup
Measurement SetupMeasurement Setup
Measurement Setup
Item Description Default value Valid Values
Temperature Temperature unit °C K; °C; °F; o
pH pH measure R R; o
mV of pH input mV of pH readings R R; o
ORP Redox measure R R; o
DO % saturation Dissolved oxygen measure R R; o
DO concentration Dissolved oxygen measure ppm ppm; mg/L; o
Conductivity Electrical conductivity Auto o; Auto; 1 µS; 0.001 mS;
measure 0.01 mS; 0.1 mS; 1 mS; Auto mS
Actual conductivity No temperature compensated Auto o; Auto; 1 µS; 0.001 mS;
conductivity measure 0.01 mS; 0.1 mS; 1 mS; Auto mS
Resistivity Resistivity measure MΩ·cm Ω·cm; kΩ·cm; Ω·cm
TDS Total dissolved solids Auto o; Auto; 1 ppm; 0.001 ppt;
measure 0.01 ppt; 0.1 ppt; 1 ppt; Auto ppt
Salinity Salinity measure R R; o
Seawater specific gravity Specific gravity measure σ
t
o; σ
t
; σ
0
; σ
15
Atmospheric pressure Atm. pressure measure o o; mmHg; inHg; mbar;
psi; atm; kPa
GPS data (*) Format of the GPS coordinates o o; xx°xx’xx.x”;
xx°xx.xxx’; xx.xxxxx°