53
7.6 ORP
From “GLP data” menu select the “ORP” option and press <OK>.
All data about last ORP calibration appear: calibrated point, time and date.
If the factory calibration was restored, the dis-
play shows the warning message “Factory calibr.
restored”.
Use the arrow keys to scroll the last 5 memo-
rized calibrations.
NotesNotes
NotesNotes
Notes If no ORP calibration has been performed, a warning message advise
the user. Press <OK> to return to the previous screen.
If the selected calibration is a factory calibration, the meter shows the
message “Factory calibration”.
7.7 TEMPERATURE
From “GLP data” menu select the “Temperature” option and press <OK>.
All information about last temperature calibra-
tion appears: calibrated point, time and date.
Use the arrow keys to scroll the last 5 memo-
rized calibrations.
NotesNotes
NotesNotes
Notes If no temperature calibration has been performed, a warning message
advise the user. Press <OK> to return to the previous screen.
If the selected calibration is a factory calibration, the meter shows the
message “Factory calibration”.