28
System SetupSystem Setup
System SetupSystem Setup
System Setup
Item Description Default value Valid values
Date Update calendar YYYY-MM-DD YYYY-MM-DD;
MM/DD/YYYY;
DD/MM/YYYY
Time Update clock hh:mm:ss (24 hours) hh:mm:ss (12 hours);
hh:mm:ss (24 hours)
Power save (min) Auto shut-off or sleeping mode 5 min NO; 5; 10; 15;
after a set period of non use 20; 30; 60 min
Log interval Period between 2 subsequent 00:00:01 00:00:01 to 03:00:00
automatic records
Ref. temperature Reference temperature for 25°C 20°C; 25°C
conductivity measurements
Temp. coefficient Temperature coefficient for 1.90%/°C 0.00 to 6.00%/°C
conductivity measurements
TDS factor Conversion factor from 0.50 0.00 to 1.00
conductivity to TDS readings
TDS unit Measure unit for TDS ppm-ppt ppm-ppt; mg/L-g/L
Average length Number of readings for 01 1 to 30
average value calculation
Key beep Acoustic signal for key pressed R R ; o
Error beep Acoustic signal for wrong R R ; o
key pressed
Decimal separator Symbol used for decimal separator . . ; ,
of displayed numbers
LCD contrast Contrast level for the LCD 8 0 to 15
Distance unit (*) Measure unit for distance m-km m-km; ft-mi
GPS unit ON (*) GPS function enabled o R ; o
GPS power save (*) GPS automatically switches off R R ; o
to save battery life
Password Password insertion - Max 25 characters
Language Interface language English English; Español; Français;
Português; Italiano
ID Meter identification code - Max 25 characters
(*) only for model featuring GPS system.