Inspect the electrode for scratches or cracks. If any is present, replace the
electrode.
Cleaning procedure: clean frequently the probe by soaking it for 1 minute in
HI70670HI70670
HI70670HI70670
HI70670 or
HI70671HI70671
HI70671HI70671
HI70671 cleaning solution. After cleaning soak the electrode in
HI70300HI70300
HI70300HI70300
HI70300 storage solution before taking measurements.
Maintenance of EC Probe
After every series of measurements, rinse the probe with tap water.
If a more thorough cleaning is required, clean the probe with the supplied
brush or a nonabrasive detergent.
Notes
After a cleaning procedure, always recalibrate the system.
Grease the O-rings before installing back the sensors.