METHOD PROCEDURES
39
39
AMMONIA LOW RANGE
11. METHOD PROCEDURES
11.1. AMMONIA LOW RANGE
SPECIFICATIONS
Range 0.00 to 3.00 mg/L (as NH
3
-N)
Resolution 0.01 mg/L
Accuracy ±0.04 mg/L ±4% of reading at 25 °C
Light Source LED with narrow band interference filter @ 420 nm
Method Adaptation of the ASTM Manual of Water and Environmental Technology,
D1426 Nessler Method
REQUIRED REAGENTS
Code Description Quantity
HI93700A-0 Ammonia Low Range Reagent A 4 drops
HI93700B-0 Ammonia Low Range Reagent B 4 drops
REAGENT SETS
HI93700-01 Reagents for 100 tests
HI93700-03 Reagents for 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
• Select the Ammonia LR method using the procedure described in the METHOD SELECTION section.
Fill the cuvette with 10 mL of unreacted sample (up to the mark).
Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and close the lid.
• Press Zero. The display will show “-0.0-” when the meter is zeroed and ready for measurement.