METHOD PROCEDURES
57
57
NITRATE
11.7. NITRATE
SPECIFICATIONS
Range 0.0 to 30.0 mg/L (as NO
3
-
- N)
Resolution 0.1 mg/L
Accuracy ±0.5 mg/L ±10% of reading at 25 °C
Light Source LED with narrow band interference filter @ 525 nm
Method Adaptation of the Cadmium Reduction Method
REQUIRED REAGENTS
Code Description Quantity
HI93728-0 Nitrate Reagent 1 packet
REAGENT SETS
HI93728-01 Reagents for 100 tests
HI93728-03 Reagents for 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
• Select the Nitrate method using the procedure described in the METHOD SELECTION section.
Fill the cuvette with 10 mL of sample (up to the mark). Replace the
plastic stopper and the cap.
• Insert the cuvette into the holder and close the lid.
• Press Zero. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
10 mL