EasyManua.ls Logo

Hanna Instruments HI83325 - Ammonia High Range

Hanna Instruments HI83325
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
45
METHOD PROCEDURES
45
45
AMMONIA HIGH RANGE
11.3. AMMONIA HIGH RANGE
SPECIFICATIONS
Range 0.0 to 100.0 mg/L (as NH
3
-N)
Resolution 0.1 mg/L
Accuracy ±0.5 mg/L ±5% of reading at 25 °C
Light Source LED with narrow band interference filter @ 420 nm
Method Adaptation of the ASTM Manual of Water and Environmental Technology,
D1426, Nessler Method
REQUIRED REAGENTS
Code Description Quantity
HI93733A-0 Ammonia High Range Reagent A 4 drops
HI93733B-0 Ammonia High Range Reagent B 9 mL
REAGENT SETS
HI93733-01 Reagents for 100 tests
HI93733-03 Reagents for 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Select the Ammonia HR method using the procedure described in
the METHOD SELECTION section.
Add 1 mL of unreacted sample to the cuvette using a 1 mL syringe.
Use the pipette to fill the cuvette up to the 10
mL mark with HI93733B-0 Ammonia High
Range Reagent B. Replace the plastic stopper
and the cap. Swirl to mix the solution.
• Insert the cuvette into the holder and close the lid.
1 mL
10 mL